Factors Affecting the Efficacy of Ureterorenoscopic Holmium: YAG Laser Lithotripsy in Ureteral Stones

Üreter Taşlarında Üreterorenoskopik Holmiyum:YAG Lazer Litotripsi Etkinliğini Etkileyen Faktörler

Author information
  1. Sağlık Bilimleri Üniversitesi, İstanbul Bakırköy Dr. Sadi Konuk Sağlık Uygulama ve Araştırma Merkezi
Issue
Volume 15 · Issue 3
Pages
155-163
Published
September 27, 2023
DOI
https://doi.org/10.54233/endouroloji.20231503-1353544 (opens in a new tab)

Abstract

Objective: Holmium:Yttrium Aluminum Garnet (YAG) laser for ureterorenoscopic lithotripsy (URS-LL) is the gold standard. The most important factors affecting laser lithotripsy are stone volume, stone density, location of the stone, laser settings and properties of the laser fiber. We aimed to measure the predictive factors affecting the efficiency of lithotripsy with objective methods.

Materials and Methods: Between October 2020-February 2022 ureterorenoscopic laser lithotripsy (URS-LL) cases performed for ureteral stones in the our hospital. It was examined retrospectively. Patients with remaining clinically significant stones were not included in the study. Holmium:YAG laser 550 μm fiber was used. Demographic data of the patients, stone size, stone volume, stone density (hounsfield unit -HU) was recorded. The energy (joule-J), frequency (hertz-Hz), power (watt) values and total lasing time of the patients determined and the total laser energy amount was calculated. After the data is obtained, the total amount of energy is divided by the stone volume and the amount of energy required to fragment 1 mm³ stone (J / mm³) was calculated. Additionally, dividing the stone volume by on total laser time, fragmented stone volume per second (mm³ / sec) was calculated.

Results: Energy, frequency and power did not differ significantly (p>0.05) between groups with laser duration ≤240 sec and >240 sec. Stone volume, stone HU value, total energy, and energy used to fragment 1 mm³ stone were significantly higher in the group with laser duration >240 sec than in the group with laser duration ≤240 sec. The fragmented stone volume in 1 second (mm³/s) in the group with laser duration >240 sec was significantly lower than the group with laser duration ≤240 sec. Significant efficacy of stone 1050 HU cut-off value was observed in the differentiation of patients with laser duration ≤240 sec and >240 sec. In the group with total energy >2750 J, stone volume, stone HU value, total energy, and energy used to break 1 mm3 stone were significantly (p<0.05) higher than in the group with total energy ≤2750 J. In the group with total energy >2750 J, the fragmented stone volume in 1 second (mm³/sec) was significantly (p<0.05) lower than in the group with total energy ≤2750 J.

Conclusion: In order to determine the effectiveness of laser lithotripsy more objective data is needed, such as fragmented stone volume in 1 second (mm³/s) and the amount of energy required to the fragment 1 mm³ stone (J/mm³).