e-ISSN 2148-0532 DOI 10.54233 Open Access · CC BY-NC-SA 4.0 Türkçe
Endoüroloji Bülteni — Endourology Bulletin Submit Manuscript
Original Article Open Access

Lesion-Based Comparison of Storz Professional Image Enhancement System and White-Light Cystoscopy for Detecting Bladder Carcinoma In Situ

Mesane Karsinoma İn Situ Tanısında Storz Professional Image Enhancement System ile Beyaz Işık Sistoskopisinin Lezyon Bazlı Karşılaştırılması

Issue
Volume 18 · Issue 1
Published
2026

Abstract

Objective: This study aimed to evaluate whether the Storz Professional Image Enhancement System (SPIES) offers superior diagnostic accuracy compared with white-light cystoscopy (WLC) in the detection of bladder carcinoma in situ (CIS), given the limited evidence available for CIS-suspicious lesions.

Material and Methods: This retrospective study included 46 patients who underwent cystoscopy under local anesthesia for microscopic or macroscopic hematuria between July 2022 and July 2025. Patients with prior cystoscopy, pelvic radiotherapy, or active urinary tract infection were excluded. CIS-suspicious lesions identified during WLC prompted scheduling for TUR-BT under general anesthesia. During TUR-BT, all lesions were evaluated sequentially with WLC followed by three SPIES modes (Clara+Chroma, Spectra A, and Spectra B) by a single experienced urologist. A total of 95 CIS-suspicious lesions were recorded, individually resected, and submitted for separate pathological analysis.

Results: A total of 95 CIS-suspicious lesions were identified in 46 patients, of which 28 were CIS-positive. Lesion-based diagnostic accuracy was 30.5% for WLC and 33.7% for SPIES. SPIES detected 23 additional lesions not visualized with WLC, corresponding to an incremental lesion detection contribution of 24.2% overall and 31.9% relative to WLC-positive findings. Among CIS-positive lesions, SPIES provided an incremental diagnostic gain of 35.7% and a 55.6% relative improvement over WLC. McNemar’s test demonstrated a statistically significant difference in paired detection outcomes between the modalities (χ² = 8.1, p = 0.004), confirming the superior detection performance of SPIES.

Conclusions: SPIES outperformed WLC in detecting bladder CIS, indicating a meaningful diagnostic benefit that warrants validation in prospective comparative studies.

More From This Issue